Giving the best gas mixture  for your cell culture, the incuMIX™ has been designed to deliver superior medical graded gas quality using advanced sensor technology that constantly surveys and adjust the environment for your cell cultures. The superior design is based on many years experience with cell culture and IVF. One of the key parameters such as the gas concentration , the temperature for the media equilibration chamber and the heating plate can be adjusted as one of the examples of a superior feature. The incuMIX™ technology offers optimal conditions for your cell culture  via closely monitored key environment parameters that gives a high quality and consistent environment.

 Model

            incuMIX™            

 Temperature  Features
 Temperature Control Range

ambient to 40°C

 Individual Temperature Control

0.1°C steps

 Temperature Uniformity

±0.1°C @ 37°C  and @ 25°C ambient temperature

 Temperature Sensors

2 x PT1000

 Heating Plate

421 x 231 mm

 Media Chamber

Gas & Temperature Controlled

 Gas mixing features
 CO2 range

0-20% CO2

 CO2 control resolution

±0.1% CO2

 CO2 sensor type

Dual channel Infrared CO2 Sensor

 CO2 consumption < 2 l /hour *
 O2 range

1-21% O2

  O2 control resolution

±0.1% O2

 O2 sensor type

Fuel Cell – Medical Certified

 N2 consumption

< 12 l/hour**

 # of gas Inputs

 Gas inputs –  100 % CO2, 100 % N2

 Input Gas Pressure

1 barg

 Output Gas Pressure

0 – 2 barg

 Buffer Tank Volume

5 l

 External logging & alarms Optional
 Web Interface  

requires static IP address only & VPN access  – Android App available

 SMS Alarm

Yes

Yes

 E-mail alarm

Yes

Yes

 Automatic relay alarm

Yes

Yes

 Touch Screen Display

5.7”

 Exterior Dimensions

22.8″ W x 22.8″ D x 9.4″ H (580 mm x 580 mm x 240 mm)

 Exterior Cabinet Construction

Powder Coated Aluminium

 # of Chambers

1 standard

 External PT1000 sensors

2 ports available for external monitoring (optional)

 External gas monitoring

1 valve controlled port available

 Electrical                                          

110 – 230 VAC 50/60 Hz 3.5A

 Shipping Weight

77 lbs. (35 kg.)

 Note!Specifications are subject to change without notice!All sensor data are measured @ sea level  and normal average atmospheric pressure. * output consumption dependent** output consumption dependent