Tips: Dear customer, if you have any questions or want to know about our products, you can click the “I want to ask” button to fill in your questions.
Unsuccessful submission? Please contact info@tech2art.com
Giving the best results for your cell culture, the incuART™ has been designed to deliver superior quality using advanced sensor technology that constantly surveys and adjust the environment for your cell cultures. The superior design is based on many years experience with cell culture and IVF. One of the key parameters such as the temperature can be adjusted individually for each chamber, as one of the examples of a superior feature. The incuART™ platform offers optimal conditions for your cell culture via closely monitored key environment parameters that gives a high quality and consistent environment.
Tips: Dear customer, if you have any questions or want to know about our products, you can click the “I want to ask” button to fill in your questions.
Unsuccessful submission? Please contact info@tech2art.com
Scientific Studies
Documents
Model |
incuART – Standard | incuART – pH1 | |
Temperature Features | |||
Temperature Control Range |
ambient to 40 °C |
||
Temperature Control |
±0.1 °C steps |
||
Temperature Uniformity |
±0.3 °C @ 37 °C and @ 25 °C ambient temperature |
||
Temperature Sensors |
6x PT 1000 |
||
pH features | |||
pH Range |
N/A |
pH 6 to 8 |
|
pH Availability | N/A chamber 6 | ||
pH measurement accuracy | 0.05 pH units (within pH 7.00 to 7.60) | ||
pH resolution | 0.01 pH units | ||
Gas mixing features | |||
CO2 range | 0-20% CO2 | ||
CO2 control resolution | ±0.1% CO2 | ||
CO2 sensor type | Dual channel Infrared CO2 Sensor | ||
CO2 consumption | < 2 l /hour | ||
O2 range | 1-21% O2 | ||
O2 control resolution | ±0.1% O2 | ||
O2 sensor type | Fuel Cell – Medical Certified | ||
N2 consumption | < 12 l/hour | ||
# of gas Inputs | Gas inputs – 100 % CO2, 100 % N2, Or premixed | ||
External logging & alarms | |||
Web Interface | – Android App available | ||
SMS Alarm * | Yes | Yes | |
E-mail alarm ** | Yes | Yes | |
Automatic relay alarm *** | Yes | Yes | |
Touch Screen Display | 7” | ||
Barcode scanner ID | N/A | 2D patient ID barcode* | |
Exterior Dimensions | 22.8″ W x 22.8″ D x 9.4″ H
(580 mm x 580 mm x 240 mm) |
||
Exterior Cabinet Construction | Powder Coated Aluminum | ||
# of Chambers | 6 standard | ||
VOC Measurement **** |
Yes Yes |