Atomic Absorption Spectrophotometer (Flame Method)

Model : SP-AA3604F; SP-AA3608F Optical System : Single Beam Monochromator : Czerny-Turner type, focal lengths at 350mm Wavelength Range (nm) : 185-900nm Wavelength Accuracy(nm)

Tips: Dear customer, if you have any questions or want to know about our products, you can click the “I want to ask” button to fill in your questions.
Unsuccessful submission? Please contact info@tech2art.com

 

 

Scientific Studies

 

Documents

 

Atomic Absorption Spectrophotometer is widely used in the analysis of major and trace elements in metallurgy, geology, mining, petroleum, light industry, agriculture, medicine, health, food and environmental monitoring.

Atomic Absorption Spectrophotometer Features:

  1. Full titanium burner, stable combustion and fast heat balance.
  2. Excellent full-band deuterium lamp background correction technology.
  3. Simple, professional, automatic and comprehensive RG-WinAAS workstation software
  4. Excellent total reflection optical system ensures excellent signal-to-noise ratio of the instrument
  5. Comprehensive and safe smart gas control module.
  6. Fully automatic instrument control, the light source supports high-performance element lamps, and can be quickly completed with one click
  7. Novelly designed atomizer and intelligent lifting.

Atomic Absorption Spectrophotometer Specifications:

Model SP-AA3604F SP-AA3608F
Optical System Optical System Single Beam
Monochromator Czerny-Turner type, focal lengths at 350mm
Grating Grating 1800 lines/mm, scintillation wavelength 250nm
Spectral Bandwidth (nm) 0.1, 0.2, 0.4, 0.7, 1.4, 2nm Automated slit selection
Wavelength Range (nm) 185-900nm
Wavelength Accuracy(nm) 0.3 nm 0.2nm
Wavelength Repeatability (nm) ≤0.1nm
Spectral Bandwidth Deviation ±0.02 nm
Lamp Stand 4 8
Element Lamp Type Cu lamp(Standard accessory)
Other element lamp is optional accessory
Photometric Performance Reading Mode Transmittance, absorbance, concentration
Photometric Range 0-125%,-0.1-3.00A
Static Baseline Drift (Cu) ±0.004A/30min ±0.003A/30min
Dynamic Baseline Drift (Cu) ±0.006A/15min ±0.005A/15min
Background Correction Deuterium lamp≥ 40x (1Abs) Deuterium lamp≥ 40x (1Abs), self-priming≥ 80x (1Abs) (optional)
Flame System Characteristic Concentration (Cu) ≤0.04ug/ml ≤0.035ug/ml
Detection Limit (Cu) ≤0.008ug/ml ≤0.006ug/ml
Repeatability (Cu) RSD≤1% RSD≤0.6%
Acetylene Flow Control Auto 12-Level
Air Assist Control Auto 4-Level
Burner lift Automatic lifting
Burner Head Titanium 100mm burner
Nebulizer High efficiency glass nebulizer
Spray Chamber Corrosion-resistant material spray chamber
Safety Protection Automatically cut off gas in case of low pressure, power failure, flame out, or burner mismatched
Data Processing Measuring Method Flame method, Hydride generation-Atomic absorption method, Flame emission method
Concentration Calculation Method Standard curve method (6 linear/non-linear fitting methods), Standard addition method, Interpolation method
Repeat Measurement Times 1-30 times. Calculate and give the average value of absorbance/concentration. standard deviation and relative standard deviation
Report Printing Parameter printing, Data result printing
Other Computer External connect
Ambient Temp 10℃-30℃
Ambient Humidity 40%-85%
Power Supply 220/110V, 50/60HZ
External Size (WxDxH)(mm) 830x650x560
Shipping Size (WxDxH)(mm) 1000x780x810
N.W./G.W. (kg) 90/110
Standard Accessory Software, Air compressor, Cu Hollow cathode lamp
Optional Accessory Computer, Printer, Hydride generator, Autosampler

 